血液透析机器报警中英对照

更新时间:2023-06-03 22:53:32 阅读: 评论:0

a-pick up tube improper position
adjust pump head to tube x xx mm press blood flow key to confirm air detector not activated press air detector key
air leakage check fluid tubes connections
attach fluid tubes press lect to abort the test
attach fluid tubes press lect to start disinf
b/acetate-pick up tube improper position
battery failure change battery or press lect
排骨清汤火锅
bicart emptied press lect to confirm
bicart holder not clod clo or check bicart holder
bicart not attached to holder attach or check bicart
blood in priming detector function check stopped
blood leak detector failure clean and press blood leak key
blood pump cover not clod clo blood pump cover
blood pump overload in fch check the pump and press lect blood pump overload press blood pump key
blood pump stop time expired start blood pump
皮鞋十大品牌bpm failure turn off bpm
chem dis fill up pha completed place pick up tube in machine chemical fill up not sufficient check supply of disinfectant
connect concentrate when time to prepare machine for treatment decalcification required xx hours since last decalcific
dialysis flow too low
dialysis fluid not ready for treatment fluid ready in xxx s
dialysis fluid ready press bypass key
disconnect venous pressure press venous key to confirm disinfection required xx days since last disinfection
extra priming volume achieved
fluid bypass time expired press bypass key
fluid path obstruction check fluid tubes and dialyzer
fluid tube nsor test failed press lect to confirm
fluid tube nsor test attach fluid tubes
fluid tube nsor test remove fluid tubes
fluid tubes in safety bypass connect fluid tubes to dialyzer
fluid tubes not in safety bypass
function check restarted plea wait
heater efficiency not sufficient for proper disinfection
heparin pump overload check heparin syringe and line heparin t to zero press heparin key to confirm
high conductivity not obtained check concentrate
high inlet water cond xx ms/cm check inlet water
high monitor temperature alarm turn off the monitor impossible time or uf tting adjust time or uf volume incorrect acidic concentrate check contrate
incorrect bicarbonate conc check concentrate
incorrect concentrate check concentrate
incorrect dialysis fluid composition
insufficient inlet pressure plea check water supply
low battery power machine turns off within one minute negative uf rate check and press uf volume
no backfiltration warning press tmp to confirm
no bpm values check bpm cuff
no cuff attached attach or check bpm cuff
only blood monitor is running press lect to start fluid part dlea wait 30 c. to bypass dialysis fluid
power failure, battery operation battery operated for xx minutes ready for venous pressure test press venous key to start recovered from power failure press time key to confirm remove fluid tubes press lect to abort the test
sags test failed
t dialysis flow not reached readjust flow or call rvice
tmp limits too wide press tmp key to adjust
to empty bicart, press lect
to end treatment, press isol ufand zero time
to obtain total uf volume isol uf must be started
to start the air detector test: fill the venous drip chamber
too high tmp required increa time or decrea uf volume treatment started no ret made press lect if ok, el retuf uf rate limit reached re-adjust time or uf volume
uf rate lower than minimum t adjust time, uf volume or min uf uf stop time expired press start/stop key
uf volume t to zero press uf volume key to confirm
venous low limit below +10 mmhg press venous key to adjust venous pressure limits too wide press venous key to adjust venous pressure test interrupted press lect to confirm wrong disinfectant agent press lect to restart disinf
A-PICK UP TUBE IMPROPER POSITION
读书随笔A管位置不对
ADJUST PUMP HEAD TO TUBE X XX MM PRESS BLOOD FLOW KEY TO CONFIRM 改动了血泵管径,压血流量键确认
AIR DETECTOR NOT ACTIVATED PRESS AIR DETECTOR KEY
空气监测没启动,压空气监测键
AIR LEAKAGE CHECK FLUID TUBES CONNECTIONS看破红尘的意思
漏气,检查透析液管路接头
ATTACH FLUID TUBES PRESS SELECT TO ABORT THE TEST
接上透析液管路压箭头终止测试
ATTACH FLUID TUBES PRESS SELECT TO START DISINF
接上透析液管路压箭头开始消毒
B/ACETATE-PICK UP TUBE IMPROPER POSITION
B管位置不对
BATTERY FAILURE CHANGE BATTERY OR PRESS SELECT
电池失效,换电池或压箭头键
BICART EMPTIED PRESS SELECT TO CONFIRM
干粉简排空,压键头确认
BICART HOLDER NOT CLOSED CLOSE OR CHECK BICART HOLDER
干粉支架未关好
BICART NOT ATTACHED TO HOLDER ATTACH OR CHECK BICART
干粉未接上支架,请接上
古诗300BLOOD IN PRIMING DETECTOR FUNCTION CHECK STOPPED
预冲探测器监测血或预冲探测器赃了
BLOOD LEAK DETECTOR FAILURE CLEAN AND PRESS BLOOD LEAK KEY
漏血测试失败,清洗并压漏血键
BLOOD PUMP COVER NOT CLOSED CLOSE BLOOD PUMP COVER
血泵门未关
BLOOD PUMP OVERLOAD IN FCH CHECK THE PUMP AND PRESS SELECT
自检时血泵卡死,检查血泵压箭头
美国著名城市
BLOOD PUMP OVERLOAD PRESS BLOOD PUMP KEY
血泵卡死,压血泵键
BLOOD PUMP STOP TIME EXPIRED START BLOOD PUMP
血泵停止时间过长,启动血泵
BPM FAILURE TURN OFF BPM
血压计坏了,关闭血压计
CHEM DIS FILL UP PHASE COMPLETED PLACE PICK UP TUBE IN MACHINE 消毒液吸满了,将吸液管放入机器内
CHEMICAL FILL UP NOT SUFFICIENT CHECK SUPPLY OF DISINFECTANT
消毒液不对,检测消毒液
CONNECT CONCENTRATE WHEN TIME TO PREPARE MACHINE FOR TREATMENT 请接透析液
DECALCIFICATION REQUIRED XX HOURS SINCE LAST DECALCIFIC
需要脱钙
DIALYSIS FLOW TOO LOW
透析液流量太小
DIALYSIS FLUID NOT READY FOR TREATMENT FLUID READY IN XXXs 透析液未准备好
DIALYSIS FLUID READY PRESS BYPASS KEY
透析液浓度好了,压旁路键
DISCONNECT VENOUS PRESSURE PRESS VENOUS KEY TO CONFIRM
取下静脉压保护器,压静脉压键确认
DISINFECTION REQUIRED XX DAYS SINCE LAST DISINFECTION
需要消毒
小姐好辣EXTRA PRIMING VOLUME ACHIEVED
预冲量到了
FLUID BYPASS TIME EXPIRED PRESS BYPASS KEY
旁路时间太长,压旁路键
FLUID PATH OBSTRUCTION CHECK FLUID TUBES AND DIALYZER
管路中有堵塞,检查管路和透析器
FLUID TUBE SENSOR TEST FAILED PRESS SELECT TO CONFIRM
旁路管位置感应器测试失败,压箭头确认
FLUID TUBE SENSOR TEST ATTACH FLUID TUBES
旁路管位置感应器,接上旁路
FLUID TUBE SENSOR TEST REMOVE FLUID TUBES
旁路管位置感应器,取下旁路
FLUID TUBES IN SAFETY BYPASS CONNECT FLUID TUBES TO DIALYZER 旁路管在机器上,接上透析器
FLUID TUBES NOT IN SAFETY BYPASS傩戏面具
旁路管不在机器上
FUNCTION CHECK RESTARTED PLEASE WAIT
机器自检重新开始
HEATER EFFICIENCY NOT SUFFICIENT FOR PROPER DISINFECTION
热消毒功率不够,检查电压和保险
HEPARIN PUMP OVERLOAD CHECK HEPARIN SYRINGE AND LINE
肝素泵过载,检查针筒到底和管路夹死
HEPARIN SET TO ZERO PRESS HEPARIN KEY TO CONFIRM
肝素设定为零,压甘素键确认
HIGH CONDUCTIVITY NOT OBTAINED CHECK CONCENTRATE
高电导达不到,检查A液
HIGH INLET WATER COND XX ms/cm CHECK INLET WATER
进水电导太高
HIGH MONITOR TEMPERATURE ALARM TURN OFF THE MONITOR
机器高温报警,关机器
IMPOSSIBLE TIME OR UF SETTING ADJUST TIME OR UF VOLUME
时间或脱水设定不对,调节时间或脱水量
INCORRECT ACIDIC CONCENTRATE CHECK CONTRATE
A液不对