学术论著
中国医学装备2021年3月第18卷第3期 China Medical Equipment 2021 March V ol.18 No.3
[文章编号] 1672-8270(2021)03-0047-04 [中图分类号] R197.39 R759.1 [文献标识码] A
The exploration of application values of the detection methods of two kinds of automatic chemiluminescence immunoassay analyzer in syphilis detection/WANG Qian, MENG Qing-han, ZHAO Jin-wen, et al//China Medical Equipment,2021,18(3):47-50.
[Abstract] Objective: T o explore the application value of two kinds of automatic chemiluminescence immunoassay analyzer (Abbott i2000 and Siemens Centaur XP) in syphilis detection. Methods: 120 rum samples were lected from 120 patients who underwent syphilis rological screening in physical examination center, outpatient rvice and hospitalization department of hospital. All chemiluminescence micro-particle immunoassay (CMIA) of Abbott i2000 automatic chemiluminescence immunoassay analyzer and chemiluminescence immunoassay (CLIA) of Siemens centaur XP automatic chemiluminescence immunoassay analyzer were ud respectively for rological detection of T reponema pallidum antibody. The results were judged as positive when the ratio of sample absorbance to threshold value (S/CO) was larger than and equal to 1.0. And the results were compared with the S/CO results of parallel determination of specific Treponema pallidum antibody by treponema pallidum particle agglutination (TPPA). Results: Among 120 subjects, 94 cas were positive as the detection of CMIA, and the positive rate was 78.33%, while 72 cas were positive as the detection of CLIA and the positive rate was 60.00%. The positive results of TPPA experiment were 74 cas, and the positive rate of that was 61.67%. The TPPA experiment was ud as the confirmed method, the nsitivity and specificity of CMIA were 97.30% and 52.17%, while tho of CLIA were 94.59% and 95.65% respectively. There was statistically significant difference in the positive rate between CMIA and TPPA (x 2=14.101, P <0.05), and there
was statistically significant difference in the coincidence rate between them (x 2=7.937, P <0.05). There was positive correlation between each automatic immune chemiluminescence methods and the TPPA experiment, respectively. Conclusion: Both the two automatic chemiluminescence immunoassay methods have higher nsitivity and specificity when S/CO value of syphilis is greater than 10.00. CLIA test is recommended to be ud for review when S/ CO value of CMIA is between 1.00 and 10.00, which can shorten turnover time of sample. TPPA test is recommended to be added for review when S / CO value of CLIA is less than 4.70, which would ensure the accuracy of detection results.
[Key words] Syphilis; Automatic immune chemiluminescence analysis; Application
[First-author’s address] Department of Clinical Laboratory, Y anda Hospital, Langfang 066251, China.
[摘要] 目的:探讨i2000和Centaur XP两种全自动化学发光免疫分析仪在梅毒检测中的应用价值。方法:选取在医院体检、门诊和住院行梅毒血清学筛查的120例患者的120份血清标本,分别使用i2000型全自动化学发光免疫分析仪化学发光微粒子免疫试验(CMIA)和Centaur XP型全自动化学发光免疫分析仪化学发光免疫试验(CLIA),对梅毒螺旋体(TP)抗体进行血清学检测。以样本吸光度与临界值的比
入蜀记翻译及原文
值(S/CO)≥1.0为阳性,并与以TP明胶凝集试验(TPPA)梅毒特异性抗体确认方法平行测定的S/CO结果进行比较。结果:在120例受检者中,CMIA检测阳性结果94例,阳性率为78.33%;CLIA检测阳性结果72例,阳性率为60.00%;TPPA检测阳性结果74例,阳性率为61.67%。以TPPA试验作为确认方法,CMIA检测的灵敏度和特异度分别为97.30%和52.17%,CLIA分别为94.59%和95.65%。CMIA与TPPA检测的阳性率、符合率比较差异均有统计学意义(x 2=14.101,x 2=7.937;P <0.05)。两种全自动免疫化学发光分析法与TPPA检测的阳性符合率均呈正相关。结论:两种全自动化学发光免疫分析检测方法在梅毒S/CO值>10.00时均具有较高的灵敏度和特异度;当CMIA的S/CO值为1.00~10.00时可选用CLIA复核,缩短标本周转时间;当CLIA的S/CO值<4.70时可附加TPPA复核,以保证检测结果的准确性。
[关键词] 梅毒;全自动免疫化学发光分析法;应用DOI: 10.3969/J.ISSN.1672-8270.2021.03.012
王 倩① 孟庆晗① 赵晋文① 丁晓娜① 刘 朋① 王欣俞①*
探讨两种全自动化学发光免疫分析仪
检测方法在梅毒检测中的应用价值
*
*基金项目:河北省卫生青年科技课题(20180899)“常规人群梅毒实验室诊断的研究及检测结果的分析”①河北燕达医院检验科 河北 廊坊 065201作者简介:王倩,女,(1989- ),本科学历,主管技师,从事临床免疫学检验工作。
梅毒是由梅毒螺旋体(treponemapallidum,TP)感染引起的慢性全身性传播疾病,其传播途径主要为性接触和血液传播,已成为全世界重要的公共卫生问题之一。近年来,梅毒的发病率不断提高,流行范围不断扩大,因此对梅毒的及时诊断显得至关重要[1]。由于TP的体外培养困难,所以临床诊断梅毒的实验室方法仍以血清学检测为主,其感染机体后,主要产
生两类抗体,分别为梅毒特异性抗体和梅毒非特异性抗体[2]。梅毒特异性抗体比非特异性抗体出现早、消失晚,一些患者可终生检出特异性抗体,因此特异性抗体检测在梅毒诊断中更具优势,在临床辅助诊断中是一项良好的实验室检测指标[3-5]。由于各种检测方法中假阴性和假阳性各占一定比例,应根据实际工作情况和临床需要,尽量选择几种方法联合检测,制定合
学术论著
中国医学装备2021年3月第18卷第3期 探讨两种全自动化学发光免疫分析仪
检测方法在梅毒检测中的应用价值*-王 倩 等
理的检测流程,提高诊断TP感染的灵敏度和特异度,减少漏诊和误诊[6-8]。
本研究以梅毒螺旋体颗粒凝集试验(treponema pallidum particle agglutination,TPPA)为梅毒特异性抗体确认方法,分析比较i2000全自动化学发光免疫
分析仪化学发光微粒子免疫试验(chemiluminescence microparticle immunoassays,CMIA)与Centaur X P 全自动化学发光分析仪化学发光免疫试验(chemiluminescence immunoassay,CLIA)的检测结果,探讨两种全自动化学发光免疫分析方法在梅毒检测中的应用价值。1 资料与方法1.1 研究资料
选取2019年1-12月河北燕达医院体检、门诊和住院行梅毒血清学筛查的120例患者,其中男性53例,女性67例;年龄0~93岁,平均年龄(58±11)岁,患者年龄多集中在42~61岁。对120例患者行梅毒血清学筛查,获得31 110份血清标本,最终纳入120例患者的血清标本120份。本研究经医院伦理委员会批准。1.2 纳入与排除标准
(1)纳入标准。①i2000型全自动化学发光免疫分析仪初筛阳性的患者标本;②Centaur XP型全自动化学发光免疫分析仪初筛阳性的患者标本;③初筛阴性但有梅毒接触史的患者标本。
(2)排除标准。①CMIA梅毒特异性抗体检测样品吸光度与临界值的比值(S/CO)<1;②CLIA梅毒特异性抗体检测S/CO<1。1.3 仪器与试剂
采用Architect i2000型全自动化学发光免疫分析仪(美国雅培公司)及配套TP抗体诊断试剂盒;Centaur XP型全自动化学发光免疫分析仪(德国西门子公司)及TP抗体检测试剂盒;TPPA梅毒抗体检测试剂(日本富士瑞必欧株式会社)。1.4 检测方法
分别采用CMIA、CLIA和TPPA对120例患者的血清标本进行平行梅毒血清学检测,操作均严格按
照相关试剂盒说明书和仪器作业指导书(standardized operation process,SOP)进行,并做室内质量控制和阴阳性对照,在室内质量控制均在控的条件下检测标本。
1.5 观察与评价指标
(1)结果判断标准。CMIA和CLIA中仪器检测
的S/CO≥1.0即为阳性,S/CO值<1.0即为阴性;TPPA中以致敏粒子孔呈现不规则凝集反应即为阳性。记录所有标本采用3种不同检测方法的结果。
(2)将C M I A 和C L I A 的不同S /C O 值分段排序,分为1.00~2.99、3.00~4.99、5.00~6.99、7.00~8.99、9.00~10.99及>10.99的6个组,分别计算与TPPA的阳性符合率。分别以CMIA和CLIA 的S/CO值绘制受试者工作特征(receiver operating characteristic,ROC)曲线,计算ROC曲线下面积(area under curve,AUC),确定两种方法对梅毒诊断效能的最佳截断值。1.6 统计学方法
采用SPSS25.0软件进行统计学分析,计数资料以百分比(%)表示,数据分析采用皮尔森卡方检验,以P <0.05为差异有统计学意义。2 结果
2.1 CLIA 、CMIA 和TPPA 检测结果
120例患者中,CMIA检测阳性结果94例,阳性率为78.33%;CLIA检测阳性结果72例,阳性率为60.00%;TPPA检测阳性结果74例,阳性率为61.67%。CMIA与TPPA检测的阳性率比较差异有统计学意义(x 2=14.101,P <0.05),CMIA与CLIA的符合率有差异,CMIA与TPPA检测的符合率差异有统计学意义(x 2=7.937,P <0.05),CLIA与TPPA检测的符合率差异无统计学意义。以TPPA为确证方法,CMIA检测的灵敏度和特异度分别为97.30%和52.17%;CLIA检测的灵敏度和特异度分别为94.59%和95.65%,见表1。
2.2 两种方法与TPPA 检测阳性符合率
(1)当S/CO值为1.00~8.99时,CMIA各组别均有TPPA呈阴性的结果,当S/CO值>9.00时,CMIA
各组别的TPPA结果均呈阳性,见表2。
表1 120例患者TPPA 、CMIA 及CLIA 检测结果的比较(%)
检测方法阳性率x 2
值P 值阴性率符合率x 2
麻永飞
值P 值TPPA 61.67(74/120)--38.3(46/120)100(120/120)--CMIA 78.33(94/120)14.1010.00021.7(26/120)80.00(96/120)7.9370.005CLIA
60.83(72/120)
0.141
0.707
40.0(48/120)
95.83(115/120)
0.07
0.791
大学计算机专业注:表中TPPA 为梅毒螺旋体颗粒凝集试验;CMIA 为化学发光微粒子免疫试验;CLIA 为化学发光免疫试验
学术论著
中国医学装备2021年3月第18卷第3期 探讨两种全自动化学发光免疫分析仪
检测方法在梅毒检测中的应用价值*-王 倩 等
表2 不同S/CO 值分段CMIA 与TPPA 的阳性符合率
S/CO值分段CMIA TPPA TPPA CMIA与TPPA阳性符合率(%)
阳性(例)阳性(例)
阴性(例)
1.00~
2.99136746.15
3.00~
差别计件工资制4.992012860.00
5.00~
6.991914573.68
7.00~
8.991311284.62
9.00~10.9911110100.00>10.9918180100.00合计
94
72
22
76.60%
注:表中CMIA 为化学发光微粒子免疫试验;TPPA 为梅毒螺旋体颗粒凝集试验
表3 不同S/CO 值分段CLIA 与TPPA 的阳性符合率
S/CO值分段CLIA TPPA TPPA CLIA与TPPA阳性符合率(%)
阳性(例)
阳性(例)
阴性(例)
1.00~
2.9932166.67
3.00~
4.9954180.00
5.00~
6.99550100.00
7.00~
8.9910100100.00
9.00~10.99660100.00>10.9945450100.00合计
74
72
2
97.29%
注:表中CLIA 为化学发光免疫试验;TPPA 为梅毒螺旋体颗粒凝集试验
(2)当S/CO值为1.00~4.99时,CLIA各组别均有TPPA呈阴性的结果,当S/CO值>5.00时,CLIA各组别的TPPA结果均呈阳性,见表3。
两种方法与TPPA检测的阳性符合率均与S/CO 值呈正相关。ROC曲线分析结果显示,当CMIA中截断值S/CO为6.08时,灵敏度为70.30%,特异度为95.70%,AUC最大为0.908,即S/CO值=6.08为CMIA的最佳诊断截断值;当CLIA中截断值S/CO为4.70时,灵敏度为91.90%,特异度为100.00%,AUC 最大为0.991,即S/CO值=4.70为CLIA的最佳诊断截断值,见图1。
和服动漫
图1 S/CO 比值受试者工作特征曲线分析
行综合分析,可对梅毒作出诊断并分为一期、二期和三期梅毒,以及潜伏梅毒和先天梅毒等,目前该
病已被列为乙类防治管理的病种[9]。TP感染后危害性大,病程长且临床表现形式复杂,早期侵犯皮肤和生殖器官,晚期会侵犯到全身各个系统、器官,症状和体征多样[10]。同时,梅毒可通过胎盘传染胎儿,引起胎儿宫内感染,导致流产、早产、死胎或胎儿先天性梅毒,严重危害下一代的健康[11-12]。在抗生素的广泛使用下,部分梅毒患者呈隐性感染,易造成漏诊和误诊,选择合适的检测方法尤为重要,利于该病的早发现、早治疗。
人类是TP的唯一宿主,未经治疗的梅毒患者感染一年内最具传染性,因此,梅毒的血清学检测在梅毒诊断中具有重要的临床指导意义[13-14]。目前,梅毒血清学检测是临床诊断梅毒的首要依据之一,机体感染TP后,TP将作为抗原激活机体免疫反应,从而引起机体产生梅毒相关抗体,由此,临床可通过梅毒血清学检测辅助诊断早期梅毒[15]。化学发光法是近年来临床使用广泛的TP检测方法,其优点是操作简便、自动化程度高以及可批量检测。TPPA将TP Nichols株的
菌体成分包被在明胶颗粒上,与血清中的TP特异性结合呈现肉眼可见的凝集反应,该法对TP特异性抗体的检测具有高度灵敏度和特异度。
3 讨论
根据流行病学史、临床表现及实验室检查等进
学术论著
[1][2][3][4]参考文献
收稿日期:2020-04-15
虞玲,隋俊,黄菁,等.不同方法检测梅毒螺旋体
抗体的效果对照研究[J].中国医学前沿杂志(电子版),2016,8(12):142-144.
邓劲,饶郴丽,杨廷富,等.化学发光免疫分析法检
测梅毒螺旋抗体的应用评价[J].国际检验医学杂志,2015(8):1041-1042.
俞辉,林勇平,向波,等.应用ROC曲线确定化学发光
法检测梅毒特异性抗体的最佳临界值[J].热带医学杂志,2016,16(12):1504-1507.
黎翠翠,谭俊青,李蔼文,等.化学发光法检测梅毒抗
体在临床筛查试验中的应用价值[J].实验与检验医学,2018,36(2):207-208,224.
Hu M,Liu N,Cai YY,et al.[Accuracy analysis and comparison of different rological detection methods in syphilis][J].Zhonghua Yi Xue Za Zhi,2017,97(36):2844-2847.
李志艳,刘平,高健,等.梅毒螺旋体抗体筛查方法的比较性研究[J].中华检验医学杂志,2012,35(12):1176-1179.跟踪英文
邵春燕,王海平,卓海龙,等.化学发光法检测输血前梅毒特异性抗体复查策略的研究[J].临床输血与检验,2015,17(1):46-48.
夏映凤.临床常见梅毒血清学检测方法的合理选择[J].检验医学与临床,2011,8(24):3042-3044.
苏连香.化学发光法检测血清抗梅毒螺旋体特异性抗体的意义探究[J].临床检验杂志(电子版),2019,8(4):251-252.
黄文彩,魏斯琪,付文金,等.应用ROC曲线分析确定化学发光微粒子法检测TP-Ab灰区[J].检验医学与临床,2019,16(20):3031-3033.
Cerqueira LRP,Monteiro DLM,Taquette SR,et al.The magnitude of syphilis:from prevalence to vertical transmis sion[J].Rev Inst Med Trop Sao Paulo,2017,59:e78.
乔秀真.全自动化学发光仪筛查梅毒螺旋体抗体结果分析[J].河北医药,2018,40(3):452-455.
贾敏利,田慧芳,王瑞军.三种常用的不同血清学诊断方法对梅毒螺旋抗体的检测效果分析[J].中国性科学,2017,26(5):72-74.
张保平,刘珊,韩艳秋.使用化学发光法检测26707例血清抗梅毒螺旋体特异性抗体以及结果假阳性率分析[J].现代检验医学杂志,2015(2):70-73.
朱小丽,张江为.化学发光免疫分析法检测梅毒抗体在早期梅毒诊断中的应用价值[J].临床医学研究与实践,2019,4(3):106-108.
殷丽丽,姜繁明,于小鸥.化学发光法和TPPA试验检测梅毒抗体的比较及阳性符合率分析[J].中国中西医结合皮肤性病学杂志,2019,18(2):116-119.
L i Z Y ,F e n g Z R ,L i u P ,e t a l.A N N A L S EXPRESS:Screening for antibodies against Treponema pallidum With Chemiluminescent Microparticle Immunoassay:Analysis of Discordant Serology Results and Its Clinical Characterization[J].Annals of Clinical Biochemistry,2015,53(Pt 5):5
玉颗珊珊下月轮88-592.
宋燕,肖君.雅培i2000梅毒螺旋体特异性抗体检测阈值及假阳性结果分析[J].海南医学,2017,28(12):1962-1964.
张薇,车玉传.雅培ARCHITECT i2000SR在梅毒实验室诊断中的临床应用和评价[J].检验医学与临床,2018,15(24):3721-3725.
[5]
[6]
[7]
[8][9]
撒娇卖萌的句子[10]
[11]
[12][13]
[14]
[15]
[16]
[17]
[18]
[19]
本研究以TPPA为梅毒特异性抗体检测的确证方法,分别对两种全自动免疫化学发光免疫分析法进行了对比分析,其结果显示,CMIA检测的灵敏度为97.30%,特异度为52.17%;CLIA检测的灵敏度为94.59%,特异度为95.65%。CMIA S/CO值为低值时的假阳性率高,尤其是在S/CO值为1.00~3.00时,与TPPA检测的符合率为46.15%;有文献报道,i2000型全自动化学发光仪CMIA S/CO值为1~10时存在假阳性[16-18]。在张薇等[19]的研究中S/CO值为1~3时,CMIA与TPPA检测的阳性符合率仅为33.0%,临床检测中存在一定数量的假阳性结果。本研究发现Centaur XP型全自动化学发光免疫分析仪CLIA 的灵敏度低于i2000型全自动化学发光免疫分析仪CMIA,但特异度高于CMIA,具有良好的临床使用价值。
由于TPPA操作繁琐且不能自动化检测,在一定程度上增加了工作人员的负担,延长了报告周转时间。本研究结果显示,CMIA和CLIA均是理想的筛查方法,以仪器厂家提供的S/CO值1.00为临界值时,CMIA的阳性符合率为76.6%,CLIA为97.29%;当S/CO值为10.00时,两者的阳性符合率均可达100%。宋燕等[18]
研究报道,当S/CO值>10.00时无需进行复查,可以直接结果审核。4 结论
本研究中两种全自动免疫化学发光分析法在梅毒S/CO值>10.00时均具有较高的灵敏度和特异度;当CMIA的S/CO值为1.00~10.00时,建议选用CLIA复核,缩短标本周转时间;当CLIA的S/CO 值<4.70时,应附加TPPA复核,以保证结果的准确性。
中国医学装备2021年3月第18卷第3期 探讨两种全自动化学发光免疫分析仪
检测方法在梅毒检测中的应用价值*-王 倩 等