Operation of the Sorvall LYNX 4000 / 6000 Centrifuge (1)
● Switch on the Centrifuge
Turn on the power switch on the right side of the centrifuge (push the switch handle towards the back of the centrifuge).
The centrifuge performs a lf-check of its software.
● Centrifuge Door
Two gas pistons open the centrifuge door.
Check the gas pistons as described below:
1. Press “Open Door” on the touchscreen on the big key on the right corner of the centrifuge.
2. If the centrifuge door opens completely, the gas pistons are functioning properly.
3. If the centrifuge door does not open, the gas pistons need to be replaced.
Door Opening
The centrifuge door can only be opened when the centrifuge is turned on.
Press “Open Door做自己作文” on the touchscreen on the big key on the right corner of the centrifuge.
Clo Centrifuge Door
Clo the centrifuge door by pressing down on it lightly in the middle or on both sides of it.
Two locks clo the centrifuge door completely.
NOTE: The door should audibly click into places. Do not slam the centrifuge door.
● Rotor Installation
The approved rotors for the Sorvall LYNX 4000 / 6000 centrifuge are listed in ction “Rotor Selection”. U only the rotors and accessories from this list in the centrifuge.
The rotors are equipped with an Auto-Lock rotor exchange.
This system is ud to automatically lock the rotor to the centrifuge spindle. The rotor does not have to be bolted on to the centrifuge spindle.
Proceed as follows:
百忙之中1. Open the door of the centrifuge and if necessary remove any dust, foreign objects or residue from the chamber.
Auto-Lock and o-ring must be clean and undamaged.
2. Place the rotor over the centrifuge spindle and let it slide slowly down the centrifuge spindle.
The rotor clicks automatically into place.
老子孔子3. Check if the rotor is properly installed by lifting it slightly on the handle. If the rotor has not been locked, place the rotor over the centrifuge spindle again.
4. Clo the rotor with the rotor lid.
5. Clo the centrifuge door.
● Entering Parameters
Acceleration / Deceleration Profiles
There are 9 curves (1-9) for acceleration and 10 for deceleration (0-9). The acceleration / deceleration profile can be lected in the main screen.
✧ Touch the desired field and window will open.
✧ Touch the number of the desired profile or move the slider be sliding your fingers over the numbers.
✧ Select “Apply” to confirm this lection for the next run.
The profile with the lowest number has the shallowest incline marked with “min”; profile number 9 is the steepest incline which is marked with “max”.
草莓的英文怎么读 Prelecting Speed / RCF-Value
✧ Press “100,000 xg”.
✧ Press 小妹妹送我的郎“rpm” or “rcf” depending on the value you would like to change. The lected function is highlighted in yellow.
✧ Enter the desired value using the numeric pad.
The digits show in quential order.降的反义词是什么
✧ Confirm your entry by pressing “Confirm”.
Run Time Prelection
✧ Press the “30:00” key in order to open the runtime lection menu.
Press “Time”, “Hold” or ∫w2dt, depending on the value you would like to change.
Enter the desired value using the numeric pad.
The digits show in quential order.
✧ Confirm your entry by pressing “Confirm”.
Prelecting the Temperature
For the Sorvall LYNX 4000 centrifuge you can prelect temperatures between -10℃ and +40℃. For the Sorvall LYNX 6000 centrifuge清廉文化墙 you can prelect temperatures between -20℃ and +40℃.
To t the temperature, proceed as follows:
✧ Press the temperature field in order to open the temperature menu.
✧ Enter the desired value using the numeric pad.
The digits show in quential order.
✧ Confirm your entry by pressing “Confirm”.
Pre-warming or Pre-cooling the Centrifuge
To pre-temp the centrifuge, proceed as follows:
U “Configuration” and “Run” to lect a stored program. The pre-temp function is fixed on program 1. Press “Load对朋友说的话” to lect programmed parameters. In the main screen you can then t the required target temperature.