Change 修改内容: | 1 Amend attachment A <Lo book of U for Analytical Balance>. 修改附件A《分析天平使用记录》。 2 Add check procedure and related form and add calibration regulation as the attachment. 增加每天进行全范围核查步骤以及相应表格,将校正规则加于附件中。 | |||
Objective/Scope 目标/范围: | 紧逼裤 To ensure u and calibration of Analytical Balance is performed according to the standard procedure 确保分析天平的使用与校正依照本程序进行。 | |||
Definition 术语定义: | N/A | |||
Applicable to 适用对象: | AC/分析化学部门 | |||
Attachments 附件: | A:Logbook of U for Analytical Balance/《分析天平使用记录》 B:Regulation for Analytical Balance Daily Calibration/《分析天平每日校准规则》 C:Regulation for Analytical Balance Daily Check/《分析天平日常核查规则》 D:Periodic Calibration Regulation for Analytical Balance/《分析天平周期性校正规则》 E:Periodic Calibration Table/《分析天平周期性校正表格》 | |||
Reference Standards 参考标准: | Chine Computation Calibrating Regulation JJG 98-90/参考中华人民共和国计量检定规程JJG 98-90 | |||
Procedure 程序: | ||||
Responsibility 执 行 人 | Step 步骤 | Action 执行内容 | |
Analyst 分析员 | 1. | Check the balance for visual clean, without sundries. Connect Analytical Balance to the power; turn it on, warm-up for about 30 minutes. 目视检查天平表面清洗,无杂物。接通电源,打开天平开关,预热约30分钟。 | |
米粉怎么做好吃又简单 | 2. | Unload the weighting pan and press “on” key. 空载秤盘,点击“On”键。 | |
3. | The analyst who us the balance first in the day should calibrate it according to <Regulation for Analytical Balance Daily Calibration> (Attachment B). 每天第一个操作天平的人参照《分析天平校准规则》(附件B)对天平进行校准。 | ||
4. | Before u, check the balance according to <Regulation for Analytical Balance Daily Check> (Attachment C)。 每次使用前,按照《分析天平日常核查规则》(附件C)对天平进行核查。 | ||
5. | Place weighing sample on the weighing pan, wait until the stability detector“o”disappears. Record the result and fill in attachment A <Logbook of U for Analytical Balance> (Attachment A). 样品称量:将样品放在秤盘正中,等待稳定状态探测符“。”消失,读取称量结果,填写《分析天平使用记录》(附件A)。 | ||
6. 二十岁的你 | When not in u, keep pressing the “Off” key until “Off” appears in the display, and the balance turns off. 天平不使用时,按住“Off”键不放直到显示屏上出现“Off”字样,再松开键,天平即关闭。 | ||
7. | Send <Logbook of U for Analytical Balance> to QA for review every month and then archive it in AC. 每月将《分析天平的使用记录》送交QA复核,复核后在AC归档保存。 Note: How long should this records be rerved depends on clients’ requirement. If the client did not require it, keep them for 5 years and inform the client before destroying the records./保存时间以客户要求为主,如客户无要求,则保存5年,销毁前通知客户。 | ||
8. | Periodic calibration for analytical balance should be performed every 6 months according to <Periodic Calibration Regulation for Analytical Balance> (Attachment D). 参照《分析天平周期性校正规则》(附件D)对分析天平每半年进行一次周期性校正。 | ||
Time 使用时间段 | Check | Ur 使用人 | U objectives 使用目的 (填KF、LOD等) | Note 备注 | ||
Theoretical amount | Actual amount | |||||
09:00- 09:05 | N/A | N/A | XXX | 校正 | Calibration Pass | |
9:30-10:00 | 5g | 5.0000g | XXX | KF | N/A | |