600 02 00Sample Queues Sample inprocess queue is online The subsystem has completed a successful homing procedure.The
subsystem is available for u.The error limits are ret.
样品架样品内处理架在线子系统成功完成复位过程。子系统可以使用。错误限制被恢复。600 02 01Sample Queues Sample inprocess queue is offline An error occurred during homing which prevented the sample inprocess
queue from recovering or the error threshold was met.
样品架样品内处理架离线复位过程中发生错误,当防止样品内处理架恢复时或遇到错误的极
限位置。
600 02 02Sample Queues Sample inprocess queue cycle failure The sample inprocess queue could not move becau it was recovering
from an error on a previous cycle
样品架样品内处理架循环失败由于在上一循环中错误被恢复,样品内处理架不能准备运行。
600 02 03Sample Queues Sample inprocess queue GIOB failure Sample Handling Sequence Controller PC board detected a parity error
on the communication bus
样品架样品内处理架GIOB失败样品处理顺序控制板在通讯总线上检查到一个奇偶校验错误
600 02 04Sample Queues Tip tray exchange failure becau no tray is
prent
样品架由于没有板,加样头板更换失败
600 02 05Sample Queues Tip tray exchange failure becau tip strap is
prent
样品架由于有板扎带,加样头板更换失败
600 02 06Sample Queues Tip tray exchange failure becau of a
subasmbly error
样品架由于部件错误,加样头板更换失败
600 02 07Sample Queues No data from sample barcode scanner for Rack
Position:*1
样品架某样品架条码扫描无数据
九湖乡Data received from the sample barcode scanner is in an invalid format 600 02 08Sample Queues Sample rack barcode format is invalid for Rack
Position:*1
样品架某样品架条码扫描格式无效从样品条码扫描接受到的数据格式无效
600 02 09Sample Queues Sample tube barcode format is invalid for
Rack:*1,Tube Position:*2.
入职感悟
样品架某样品架某试管位条码扫描格式无效
我好难
600 02 10Sample Queues Sample exit queue is full
济南社保代缴样品架样品退出架列满
600 02 11Sample Queues Sample exit queue is not full
样品架样品退出架列未满
方文杰600 02 12Sample Queues Sample loader is online The subsystem has completed a successful homing procedure.The
subsystem is available for u.The error limits are ret.
样品架样品装载器在线子系统成功完成复位过程。子系统可以使用。错误限制被恢复。600 02 13Sample Queues Sample loader is offline An error occurred during homing which prevented the sample loader from
recovering or the error threshold was met.
样品架样品装载器离线复位过程中发生错误,当防止样品装载器恢复时或遇到错误的极限
位置。
600 02 14Sample Queues Sample exit pusher is online The subsystem has completed a successful homing procedure.The
三亚海口
subsystem is available for u.The error limits are ret.
样品架样品架退出推动器在线子系统成功完成复位过程。子系统可以使用。错误限制被恢复。600 02 15Sample Queues Sample exit pusher is offline An error occurred during homing which prevented the sample exit pusher
from recovering or the error threshold was met.协议的英文
样品架样品架退出推动器离线复位过程中发生错误,当防止样品架退出推动器恢复时或遇到错误
的极限位置。
600 02 16Sample Queues Stat sample pusher is online The subsystem has completed a successful homing procedure.The
饭后茶余subsystem is available for u.The error limits are ret.
样品架急诊样品推动器在线子系统成功完成复位过程。子系统可以使用。错误限制被恢复。600 02 17Sample Queues Stat sample pusher is offline An error occurred during homing which prevented the stat sample pusher
from recovering or the error threshold was met.
样品架急诊样品推动器离线复位过程中发生错误,当防止急诊样品推动器恢复时或遇到错误的
极限位置。