PhoCheck TIGER SELECT Instrument 说明书

更新时间:2023-07-16 08:22:58 阅读: 评论:0

PhoCheck
TIGER SELECT
Instrument Ur Manual V1.1
rt Number: 861489
About this Manual
Contents Page
About this Manual    1
Contents    2
Statements    3
纠缠是什么意思Responsibility for u    3
Caution    3
Quality Assurance    3
Disposal    3
Calibration Facility    3
Ion Science Contact Details    3
Introduction to PhoCheck TIGER Select    4
Getting Started    5
Understanding the Display Screen    6
abac式的成语Start up screens    6
Soft Key areas    6
Using your TIGER Select 7
TIGER Select mode 8-9
TIGER PC Software 10
Maintenance 11-12
Calibration 11
Probes 12
TIGER SELECT Parts 13-14
进出口经营权Main Asmbly 13
带海字的网名
Front End Filter Asmbly 14
Accessories Plea refer to Tiger Manual Instrument Warranty and Service Plea refer to Tiger Manual Technical Specification  15
Manual History 16
Statements
RESPONSIBILITY FOR USE
Inadequate performance of the gas detection equipment described in this manual may not necessarily be lf-evident and conquently equipment must be regularly inspected and maintained.  Ion Science recommends that personnel responsible for equipment u institute a regime of regular checks to
ensure it performs within calibration limits, and that a record be maintained which logs calibration check data.  The equipment should be ud in accordance with this manual, and in compliance with local
safety standards.
LEGAL NOTICE
Whilst every attempt is made to ensure the accuracy of the information contained in this manual, Ion Science accepts no liability for errors or omissions, or any conquences deriving from the u of information contained herein.  It is provided "as is" and without any reprentation, term, condition or warranty of any kind, either express or implied. To the extent permitted by law, Ion Science shall not be liable to any person or entity for any loss or damage which may ari from the u of this manual.  We rerve the right at any time and without any notice to remove, amend or vary any of the content which appears herein.
!CAUTION!
It is esntial that the PhoCheck TIGER is always ud with a supplied PTFE 0.5 micron filter fitted to the front of the instrument.  Without a filter, particles of debris and dust can be drawn into the detector inhibiting the function of the instrument.  The filters are consumable and should be changed after every 100 hours of u.  The frequency of replacement should be incread for dusty or moisture laden environments.  Filters are available from your distributor or at .
Quality Assurance
PhoCheck TIGER has been manufactured in compliance with ISO9001:2000, which ensures that the equipment supplied to our customers has been designed and asmbled reproducibly, from traceable components, and leaves Ion Science calibrated to stated standards.
Disposal
Dispo of PhoCheck TIGER, its components and any ud batteries in accordance with all local and national safety and environmental requirements.  This includes the European WEEE (Waste Electrical and Electronic Equipment) directive.  Ion Science Ltd offers a take back rvice.  Plea contact us for more information.  The PhoCheck TIGER field ca material is recyclable polypropylene.
Calibration Facility
Ion Science Ltd offers a calibration rvice including the issue of certification confirming calibration with equipment traceable to national standards.  A PhoCheck TIGER calibration kit is available from your distributor or rvice centre or at .  Ion Science recommends annual return of all instruments for yearly rvice and calibration.
喝酒祝福语大全简短
Introduction to PhoCheck TIGER Select
Benzene vapour is carcinogenic. Benzene is a constituent of certain types of fuel, and ud as a solvent, and in the production of a wide range of chemicals and plastics.
PID technology can easily detect Benzene vapour as well as many of other volatile organic compounds.
如何做羊肉The Tiger Select us a combination of a ultraviolet and chemical filtering to deliver a respon to Benzene free from interferents.
When the Select upgrade is activated, data logging, H&S mode and ppb automatically become available.
At start up a specific Select screen appears however the instrument initially runs like a standard Tiger. When in Select mode the instrument takes control over many of the standard features.
The Tiger Select us a standard 10.6 eV lamp and a lective 10 eV filter incorporated in the
pellet/stack.
The lower eV output prevents nsing many of the gas co-prent in Benzene containing atmospheres.
10.0eV 10.6eV
蹬三轮
超广角镜
The Tiger Select starts up in normal running mode.
Ion Science recommend operating the instrument in the working environment for 10 to 15 minutes before u. This allows the tubes and instrument to acclimatize and ensures the best results.
Ion Science suggest that the TIGER Select is calibrated before each u and a new PID-Pre –Filter Tube is ud.
With no tube fitted, the customer can check the general VOC level, if there are no VOC's prent then no tube test is required.