DICOM Conformance Statement for Philips DICOM Viewer 3.0 SP3

更新时间:2023-06-18 04:16:12 阅读: 评论:0

© Koninklijke Philips Electronics N.V. 2013
All rights are rerved. Document Number: PIIOffc.0001703
DICOM Conformance Statement Philips DICOM Viewer Relea 3.0 SP3
Issued by:
Philips Healthcare
P.O. Box 10.000
5680 DA Best
The Netherlands
Email:   
Internet: www./main/about/connectivity Document Number: PIIOffc.0001703
Date:                      02-September-2013
1. DICOM Conformance Statement Overview
The Philips DICOM Viewer R3.0 is an integrated component, which means that much of its functionality is implemented by various PII software components which are re-ud to build the Philip
s DICOM Viewer for viewing DICOM data. The application consists of a ries lector and a ries viewer and supports basic image viewing operations like playing movies and adjusting image ttings. The Philips DICOM Viewer is a read-only application, i.e., changes made to the images cannot be saved, but they can be exported to the system clipboard and/or an office (i.e., paper) printer.
The Philips DICOM Viewer is a basic viewer for DICOM data. The DICOM data to be viewed can be prent (i.e., referred to) in a DICOMDIR, or as a collection of individual DICOM files, which may be stored at any Windows file system location. The Philips DICOM Viewer can be run either from hard disk or from a removable medium such as a CD, a DVD, or a USB device.
The Philips DICOM Viewer does not support DICOM printing nor is able to print to P2P film printers.
When DICOM image information will be distributed by means of CD or DVD, the DICOM viewer can be distributed together with the DICOM image information by copying the viewer onto that CD or DVD.
After inrting the CD/DVD into a system the application may be started up to allow the ur to lect one of the patient folders on the CD/DVD.
Supported Operating System (OS) are XP (Professional or Home Edition) or VISTA (Home Basic, Home Premium, Business, Enterpri or Ultimate) or WINDOWS 7 (Home Premium, Professional or Ultimate) or Windows 8.
The Philips DICOM Viewer is not intended for official clinical and/or diagnostic purpos, but only for non-clinical and non-diagnostic illustration purpos.
The Philips DICOM Viewer will, for the time being, be delivered with US English ur interface texts.
The system supports correct viewing image types simultaneously of different Multi-modality studies.
On-line help and Off-line help (including ur documentation) will not be part of this relea.
Ejection of the removable medium:
The system shall clear its patient list, pictorial index, image viewer and DICOM information panel when the removable medium containing the current DICOMDIR is ejected from the OS system.
The operator does not need any training specifically for using the Philips DICOM Viewer.
NM Layouts for Philips DICOM Viewer
Bad on the data that is loaded the system will automatically activate the preferred layout. The ur will have the cho to change the layout from a drop down list of all other available layouts. The NM Layouts are only available with a NM licen. Refer to ction 7, where this licensing mechanism is explained.
Following is the list of all supported layouts:
1. DISPLAY WB – Dual Intensity Compare
The same anterior and posterior image is prented twice - with different W/L intensities.
2. DISPLAY WB + STATICS
In this layout, the number of static views displayed is flexible, depending on the number of static images available, and can be 4 (2rx2c), 6(3rx2c) or 8 (4rx2c).
3. DISPLAY DYNAMIC + STATICS
In this layout, the number of static views displayed is flexible, depending on the number of static images available, and can be 4 (2rx2c), 6(3rx2c) or 8 (4rx2c).
4. DISPLAY WB + VOLUME
This layout prents planar Whole Body and transver, sagittal and coronal slices from a volume.
5. DISPLAY PLANAR + VOLUME
This layout prents planar static images and transver, sagittal and coronal slices from a volume.
© Koninklijke Philips Electronics N.V.  02-September-2013
Up to 4 static images may be displayed.
6. SINGLE SERIES CORONAL LAYOUT
Displays 3x4 volume tilting with image orientation as “Coronal Front”.
7. SERIES TSC + MIP – ANATOMICAL AND FUNCTIONAL LAYOUT DEFINITION
心理学测试Two ries, all 3 orthogonal views (linked), with ability to turn on/off fusion on either ries, including rotating MIP.
8. TSC + FUSION 3 ROW – ANATOMICAL AND FUNCTIONAL LAYOUT DEFINITION
Despite being able to turn fusion on / off in the above displays, an additional layout of this kind is required for the cas
where viewers wish to view all 3 planes for anatomical, functional and fusion.
9. TSC + FUSION
10. TSC + FUSION RENDERING
11. SINGLE SERIES N ON 1 – SINGLE SERIES LAYOUT DEFINITION
Display one or multiple images from 1 orthogonal view (with menu option to change to different orthogonal view and with
option to display functional, anatomical or fud as available).
12. SINGLE SERIES TSC+MIP – SINGLE SERIES LAYOUT DEFINITION
This layout will apply for any single ries. If the MIP image is not desirable (e.g. in anatomical review) it can be removed
from the layout and the axial image resized.
13. SINGLE SERIES TSC – SINGLE SERIES LAYOUT DEFINITION
This review is required to duplicate a typical review that is ud by the CT viewer. It applies to any single ries.
14. 3 SERIES – SERIES LAYOUT DEFINITION
15. 2D – SERIES REVIEW LAYOUT DEFINITION
The 2D layout is a specific u layout intended for display of Secondary Capture images and display of data from disk as
original slices.
16. DYNAMIC TSC + CINE – SERIES REVIEW LAYOUT DEFINITION
3D Cine will be activated by default.
17. SINGLE MONITOR COMPARISON LAYOUT SERIAL
18. SLAB ANATOMICAL TSC + VOLUME LAYOUT
The Slab Anatomical TSC + Volume layout us slab viewers and prents the anatomic study when fusion is inactive.
19. SLAB TSC + FUSION 3 ROW LAYOUT
The Slab TSC + Fusion 3 row layout us slab viewers. The anatomical study is ud to t the thickness and orientation.
20. STRESS REST THREE VIEW – CARDIAC LAYOUT DEFINITION
This is a layout specific for Cardiac images.
21. STRESS REST RAW COMPOSITE VIEW
This is a Cardiac specific layout, which provides the comparison of SA, VLA and HLA Stress/Rest images along with the raw  projection image obtained from the Acquisition system.
SA stands for Short Axis, VLA stands for Vertical Long Axis and HLA stands for Horizontal Long Axis.
Following table shows an overview of all the supported DICOM and Private SOP Class.
Table 1: All supported Image SOP Class
SOP Class
Name UID
Computed Radiography (CR) Image Storage    1.2.840.10008.5.1.4.1.1.1
Digital X-Ray (DX) Image Storage – for Prentation    1.2.840.10008.5.1.4.1.1.1.1
Digital X-Ray (DX) Image Storage – for Processing    1.2.840.10008.5.1.4.1.1.1.1.1
Digital Mammography X-Ray (MG) Image Storage – for Prentation    1.2.840.10008.5.1.4.1.1.1.2
Digital Mammography X-Ray (MG) Image Storage – for Processing    1.2.840.10008.5.1.4.1.1.1.2.1
Computed Tomography (CT) Image Storage    1.2.840.10008.5.1.4.1.1.2
Enhanced Computed Tomography (Enhanced CT) images Storage.    1.2.840.10008.5.1.4.1.1.2.1
Ultrasound Multi-frame Image Storage (Retired)    1.2.840.10008.5.1.4.1.1.3
Ultrasound Multi-Frame Image (USMF) Storage    1.2.840.10008.5.1.4.1.1.3.1
Magnetic Resonance (MR) Image Storage    1.2.840.10008.5.1.4.1.1.4
Enhanced MR Image  (Enhanced MR) Storage    1.2.840.10008.5.1.4.1.1.4.1
Nuclear Medicine Image Storage (Retired)    1.2.840.10008.5.1.4.1.1.5
心里不是滋味
Ultrasound Image Storage (Retired)    1.2.840.10008.5.1.4.1.1.6
简单的水彩画© Koninklijke Philips Electronics N.V.  02-September-2013
SOP Class
Name UID Ultrasound Image (US) Storage    1.2.840.10008.5.1.4.1.1.6.1 Secondary Capture Image (SC) Storage    1.2.840.10008.5.1.4.1.1.7 Multi-frame Grayscale Byte Secondary Capture Image Storage    1.2.840.10008.5.1.4.1.1.7.2 Multi-frame Grayscale Word Secondary Capture Image Storage    1.2.840.10008.5.1.4.1.1.7.3 Multi-frame True Color Secondary Capture Image Storage    1.2.840.10008.5.1.4.1.1.7.4
X-Ray Angiographic Image (XA) Storage    1.2.840.10008.5.1.4.1.1.12.1 Enhanced X-Ray Angiographic (Enhanced XA) images Storage.    1.2.840.10008.5.1.4.1.1.12.1.1 X-Ray Radiofluoroscopic Image Storage (RF)    1.2.840.10008.5.1.4.1.1.12.2 Enhanced X-Ray Radiofluoroscopic (Enhanced XRF) images Storage.    1.2.840.10008.5.1.4.1.1.12.2.1 Breast Tomosynthesis Image Storage    1.2.840.10008.5.1.4.1.1.13.1.3 Nuclear Medicine Image (NM) Storage    1.2.840.10008.5.1.4.1.1.20 Positron Emission Tomography Image (PET, PT) Storage    1.2.840.10008.5.1.4.1.1.128 Radiotherapy Image (RT) Storage    1.2.840.10008.5.1.4.1.1.481.1 Philips Private X-Ray Image Storage    1.3.46.670589.2.3.1.1
Note that the PMS DICOM Viewer shall ignore unsupported DICOM SOP class.
Following table shows an overview of the supported DICOM and Private Prentation State SOP Class.
Table 2: Supported Prentation State SOP Class
SOP Class
Name UID Grayscale Softcopy Prentation State (GSPS, PS) Storage    1.2.840.10008.5.1.4.1.1.11.1
Philips Private Grayscale Softcopy Prentation State Storage    1.3.46.670589.2.2.1.1
The DICOM Viewer shall apply (one of) the Prentation State(s) that is provided with the image data that is displayed.
Following table shows an overview of the supported DICOM Media Services.
Table 3: DICOM Media Services
Media Storage Application Profile Write Files
FSC
Write Files
FSU
Read Files
FSR
Compact Disk – Recordable
Basic Cardiac X-Ray Angiographic Studies on CD-R  No No Yes
General Purpo CD-R Interchange No No Yes
CT/MR Studies on CD-R  No No Yes
DVD
Basic Cardiac X-Ray Angiographic Studies on DVD No No Yes
General Purpo DVD Interchange No No Yes
CT/MR Studies on DVD Media No No Yes
USB
鱼腥草的功效与作用及禁忌General Purpo USB Media Interchange with JPEG No No Yes
Note that the Philips DICOM Viewer supports the following DICOM media: CD-R, CD-RW, DVD+R, DVD-R, DVD+RW and DVD-RW.
© Koninklijke Philips Electronics N.V.  02-September-2013
2. Table of Contents
1.DICOM CONFORMANCE STATEMENT OVERVIEW (3)
2.TABLE OF CONTENTS (6)
3.INTRODUCTION (8)
3.1.REVISION HISTORY (8)
3.2.AUDIENCE (8)
3.3.REMARKS (8)
3.4.DEFINITIONS, TERMS AND ABBREVIATIONS (9)
3.5.REFERENCES (10)
结婚的英语4WORKING (11)
5.MEDIA INTERCHANGE (12)
5.1.IMPLEMENTATION MODEL (12)
5.1.1.Application Data Flow Diagram (12)
各种茶的功效5.1.2.Functional Definitions of AE's (13)
5.1.3.Sequencing of Real World Activities (13)
5.1.4.File Meta Information for Implementation Class and Version (13)
5.2.AE SPECIFICATIONS (14)
5.2.1.Philips DICOM Viewer - Specification (14)
5.2.1.1.File Meta Information for the Philips DICOM Viewer (14)
5.2.1.2.Real-World Activities (15)
左怎么读
5.2.1.2.1.Display Directory (15)
5.3.AUGMENTED AND PRIVATE APPLICATION PROFILES (19)
5.3.1.Augmented Application Profiles (19)
5.3.1.1.AUG-XABC-CD (19)
5.3.1.1.1.SOP Class Augmentations (19)
5.3.1.1.2.Directory Augmentations (19)
5.3.1.1.3.Other Augmentations (19)
5.3.1.2.AUG-GEN-CD (19)
5.3.1.2.1.SOP Class Augmentations (19)
5.3.1.2.2.Directory Augmentations (19)
5.3.1.2.3.Other Augmentations (19)
5.3.1.3.AUG-CTMR-CD (19)
5.3.1.3.1.SOP Class Augmentations (19)
5.3.1.3.2.Directory Augmentations (19)
5.3.1.3.3.Other Augmentations (19)
5.3.1.4.AUG-CTMR-CD (20)
5.3.1.4.1.SOP Class Augmentations (20)
5.3.1.4.2.Directory Augmentations (20)
5.3.1.4.3.Other Augmentations (20)
5.3.2.Private Application Profiles (20)
5.4.MEDIA CONFIGURATION (20)
6.SUPPORT OF CHARACTER SETS (21)
7.SECURITY (23)
7.1.SECURITY PROFILES (23)
7.2.ASSOCIATION LEVEL SECURITY (23)
7.3.APPLICATION LEVEL SECURITY (23)
8.ANNEXES (24)
8.1.IOD CONTENTS (24)
8.1.1.Created SOP Instance (24)
8.1.2.Usage of Attributes from Received IOD (24)
8.1.3.Attribute Mapping (24)
8.1.4.Coerced/Modified fields (24)
攀谈的近义词
© Koninklijke Philips Electronics N.V.  02-September-2013

本文发布于:2023-06-18 04:16:12,感谢您对本站的认可!

本文链接:https://www.wtabcd.cn/fanwen/fan/89/1043480.html

版权声明:本站内容均来自互联网,仅供演示用,请勿用于商业和其他非法用途。如果侵犯了您的权益请与我们联系,我们将在24小时内删除。

标签:功效   作用   测试   近义词   心理学   攀谈   鱼腥草   禁忌
相关文章
留言与评论(共有 0 条评论)
   
验证码:
推荐文章
排行榜
Copyright ©2019-2022 Comsenz Inc.Powered by © 专利检索| 网站地图