Figure 1. Digital Static Locator Item #19430 Desco’s Digital Static Locator is a compact
4. Hold the fieldmeter 1” away from the charged
surface. Two beams of light will form concentric
circles when the meter is the correct distance, 1”
away, from the surface. The measurement
accuracy is dependent on a stable ground reference and the 1 inch measuring distance. It is also
dependent on the “aspect ratio”, relating the size of the object to be measured to the measurement
distance. This aspect ratio should be at least 3 for
best accuracy, i.e. the object should be at least a
3 inch square when measuring at a 1 inch distance. Accurate measurements may be made at other
measurement distances by scaling the meter range and obrving the proper aspect ratio. For exam
ple, at a measurement distance of 3 inches, multiply the meter reading by 3 to give a range of 0 to 60 kilovolts. For accuracy, the object being measured at this
distance should be at least a 9 inch square.
5. A bar graph and a digital display on the 19430
汤圆由来displays the voltage of the charge. Press HOLD if you want to hold the display of the measured charge.
6. To turn off meter: Press POWER if you have not
ud the HOLD mode.
If you have presd HOLD and
wish to retain the
meter reading for the next u,
press the POWER button to
turn off the 19430.
If you have presd HOLD and
do not wish to retain the meter
reading for next u, press
HOLD then POWER.
If the 19430 is left on for five minutes, the unit will beep five times then turn itlf off automatically. To disable this feature, simply press and hold the red power button for at least three conds when turning the 19430 on. Three beeps and the A-OFF display indicates that the feature has been disabl
ed.
Results
A display of [+] or [-] next to the digital numeric value reprents the polarity of the static charge being measured. In addition, the color of the bars in the bar graph indicates the polarity of the static charge. Red and blue bars indicate positive and negative charges respectively.
During a static charge measurement, if the digital display starts flashing along with a continuous beep, stop taking the measurement. The static charge is greater than the meter can measure. Attempts to measure too high a charge may damage the nsor.Operation For Ion Balance (Offt Voltage) Measurement
NOTE: The Digital Static Locator Ion Balance measurement feature should not be ud to measure the ion balance of static bars. It may be ud to mea-sure other ionizers for ion balance (offt voltage).
1. Depress the plate asmbly relea button on either side of the fieldmeter to remove the charge plate
asmbly.
2. Inrt the plate asmbly into the other end of the
fieldmeter (the end with the “Static Locator”) until it
clicks into place.
NOTE: Ion Balance plate asmbly must not move during measurement.
3. The conductive ca must be grounded using the
supplied ground cord. It is imperative for accurate ion balance and to drain any charge that has
developed prior to measurement.
4. To turn on the ion balance measuring circuit, press the blue IB button.
5. Turn nsor toward an uncharged area.
6. The meter reading should be zero (0). Press the
Zero button if it is not.
7. Hold the fieldmeter centered in front of the ionizer at any convenient, repeatable distance, such as 1 foot.
8. A bar graph and a digital display on the 19430
displays the ion balance information. Press HOLD if you want to hold the display of the measured
charge.
9. To turn off the meter:
Press POWER if you have not ud the HOLD
mode.
If you have presd HOLD and wish to retain the
meter reading for the next u, press the POWER button to turn off the 19430.
If you have presd HOLD and do not wish to
retain the meter reading for next u, press HOLD then POWER.
10. Depress the charge plate asmbly relea buttons on either side of the fieldmeter to remove the charge plate asmbly. Store the plate on the opposite end, pressing it until it clicks into place.
Figure 3.
If the 19430 is left on for five minutes, the unit will beep five times then turn itlf off automatically. This feature can be disabled.
Troubleshooting Guide
Fault Possible
Caus
Solution
Fieldmeter is on but when turned toward a charges object, the number and the bar graph do not change.The meter is in HOLD
mode of operation.
Cancel the HOLD
mode by pressing the
上学之路
HOLD button once.
The display cannot be t to zero.The meter is in HOLD
mode.
The true ZERO point
has shifted.
Cancel the HOLD
mode by pressing the
HOLD button once.
If the meter reads
0.20 or more for an
uncharged object, a
read adjustment
may be needed.
Contact your local
reprentative or
Desco Customer
Service for
adjustment, repair or喝咖啡会长胖吗
replacement.
When the POWER button is presd, LCD display does not appear.No battery, low
battery capacity or
bad contact.
Malfunction of the
LCD display.
Replace the battery /
Make proper
connection.
Discontinue u and
contact your local
reprentative
or Desco Customer什么样的风景
Service for
adjustment, repair or
replacement.
A part of the LCD display does not appear.
Display shows ERROR.
Alarm does not sound.Malfunction of the
LCD display.
Discontinue u and
contact your local
reprentative or
Desco Customer
Service for
adjustment, repair or
replacement.
LCD display
renewal rate: 5 times/s
Accuracy: 20KV ± 10 %
Ambient conditions: 50º-100ºF, (10º - 40ºC); 0 - 60
%RH (non-condensing)
Display features: Bar graph; red LCD positive熠熠生辉
polarity voltage, blue LCD
negative polarity voltage
Bar graph
precision: ±0.1 kV for low range ±1.5 kV for
high (Hi) range; ±15 V for ion
balance range
Digital reading: Auto-ranging, three digits . = 0 to
±1.49 kV . = ±1.0 kV to
±20.0kV (Hi) = 0 to 200 (Ion
4460
Balance)
Ion Balance (offt voltage) mode display: [IB] is
displayed
Battery capacity display indicator
HOLD switch to retain display after measurement
Error sign [Error] appears if the nsor is damaged
Alarm feature: Beep sound will be heard during
the following actions:
Power on: one beep
Power on with Auto Off feature disabled: [Depress
power button for more than 3 conds] three beeps
Auto power-off: short beeps at 1 cond interval for 5
conds before power off乘方的定义
Over range: Continuous sound
Auto power-off: Power turns off automatically
after five minutes. [A.OFF] is
displayed whenever this feature
is disabled.
Power source: 9 V, 6F22Y mangane battery.
Battery life: In excess of 30 hours.
Size: 4 7/8” (123mm) (L) × 2 7/8”
(73mm) (W) × 1” (25mm)(D)
Weight: 6 oz. (170g)
Ca material: Conductive resin (ABS)
Specifications
Measuring range: 0 to ±1.49 kV (Lo range)
± 1.0 kV to ± 20.0 kV (Hi range)
(Range hysteresis: ± 1.0 kV to
± 1.5 kV)
0 to ± 200 V (Ion Balance
measuring range)
Measuring distance: 1"±1/64" (25mm ± 0.5mm); LEDs
guide for correct distance
(between charged surface and
fieldmeter)
Respon time: <1 s
Fully Charged Empty
Figure 2. Residual battery charge as displayed. NOTE: The dark area in the battery indicator display is