显微镜学术语(一)

更新时间:2023-05-20 02:33:42 阅读: 评论:0

Microscope Glossary
Below you will find explanations for many of the common words ud in microscopy.
Abbe Condenr:  A specially designed lens that mounts under the stage and is usually movable in the vertical direction.  It has an iris type aperture to control the diameter of the beam of light entering the lens system.  By changing the size of the iris and moving the lens toward or away from the stage, the diameter and focal point of the cone of light that goes through the specimen can be controlled.  Abbe condenrs really become uful at magnifications above 400X.  The condenr lens system should have a numerical aperture equal to or greater than the N.A. of the objective lens being ud.  All of our microscopes that go to 1000X u Abbe condenrs with a 1.25 N.A.  There are two types.  One is a spiral type that you turn to move it up or down and the other is on a rack and pinion system and controlled with a condenr focusing knob.  购房新政策
Achromatic lens:  计算机考试二级When light goes through a prism or lens, it is bent or refracted.  Some colors refract more than others and as a result, will focus at different points, reducing resolution.  To help correct this problem, achromatic lens are ud.  The lens are made of different types of glass with different indexes of refraction.  The result is a better (but not perfect) alignment of some of the colors at the focal point, thereby giving you a clearer image. 
Arm:  The part of the microscope that connects the tube to the ba.  When carrying a microscope, grab the arm with one hand and place your other hand under the ba.
Articulated Arm:  A type of stand that holds a microscope body.  The stand clamps to a table and has a variety of motion in three dimensions. 
Ba:  The bottom support of the microscope (e arm above).
Binocular Head:  A microscope head with two eyepiece lens, one for each eye.  Generally this term is ud in describing a high power (compound) microscope.  With a low power microscope we say "stereo" head becau, unlike the compound microscope, the stereo has a parate objective lens for each eyepeice lens, producing two independent paths of light, one for each eye.  In the compound microscope with a binocular head, there are two eyepiece lens but still only one objective lens and you will not get stereo vision.
Body:  This term is ud mostly with the low power stereo microscopes and it is the basic heart of the microscope without any type of stand (ba) or illuminators.  It usually includes the eyepiece and objective lens but not the focusing block.
C-mount:  This is an adapter ud with various types of video cameras.  Usually, you unscrew the lens from the camera and screw in the adapter.  The adapter then connects to the trinocular port on the microscope.
Coar Focus:  This is the rough focus knob on the microscope.  You u it to move the objective lens toward or away from the specimen (e fine focus).
Coaxial Focus:  A focusing system that has both the coar and fine focusing knobs mounted on the same axis.  Usually the coar knob is larger and on the outside and the fine knob is smaller and on the inside.  On some coaxial systems, the fine adjustment is calibrated, allowing differential measurements to be recorded.
Condenr Lens:  A lens mounted in or below the stage who purpo is to focus or conden the light onto the specimen.  The higher power objective lens have very tiny diameters and require concentrated light to work properly.  By using a condenr lens you will increa the Illumination and resolution.  Condenr lens are not required on low power microscopes.
Contrast Plate:  A circular opaque plate placed on the stage of a low power microscope.  One side is white, the other is black.  It can be flipped around depending on the coloration of your specimen.
Cover Slip:  A very thin square piece of glass or plastic placed over the specimen on a microscope slide.  When ud with liquid samples, it flattens out the liquid and assists with single plane focusing.
Diaphragm:  Generally a five hole disc placed under the stage on a high power microscope.  Each hole is of a different diameter.  By turning it, you can vary the amount of light passing through the stage opening.  This will help to properly illuminate the specimen and increa contrast and resolution.  The diaphragm is most uful at the higher powers.
DIN Optics:  A German standard for the manufacturing of microscope lens.  DIN lens aren't particularly better than non-DIN but they will be interchangeable from one DIN microscope to another.  They are t to work with a 160mm tube length and have a uniform thread.  Most quality microscopes u DIN optics.
Diopter Adjustment:  回的反义词When you look through a microscope with two eyepiece lens, you must be able to change the focus on one eyepiece to compensate for the difference in vision between your two eyes.  The diopter adjustment does this.  The way to correctly adjust this is to first clo the eye over the eyepiece with the diopter adjustment and normally focus the microscope so that the open eye es the image in focus.  Next, switch eyes (clo the open eye, open the clod eye) and without changing the main focus knobs, focus on the image by turning the diopter adjustment only.  Now with both eyes open, the image should be clear with both eyes.  (This technique is ud with binoculars too.)
Dual Head:  A microscope (usually high power) with a single eyepiece lens coming out one side and an additional single eyepiece tube coming either off the top or from the opposite side.  Dual heads are ud so that a teacher can verify what a student is eing or can be ud for video or camera work.  It is not recommended that two students do a lab sharing a single dual microscope as it will get to be uncomfortable for the student using the top eyepiece.
Eyepiece Lens:  The lens at the top of the microscope that you look into.  They are usually 10X but also are available in 5X, 15X and 20X.  Widefield lens have a large diameter and show a wide area of the field of view. 
Fine Focus:  This is the knob ud to fine tune the focus on the specimen.  It is also ud to focus on various parts of the specimen.  Generally one us the coar focus first to get clo then moves to the fine focus knob for fine tuning. 
Field of View:  Sometimes abbreviated "FOV", it is the diameter of the circle of light that you e when looking into a microscope.  As the power gets greater, the field of view gets smaller.  You can measure this by placing a clear metric ruler on the stage and counting the millimeters from one side to the other.  Typically you will e about 4.5mm at 40X, 1.8mm at 100X, 0.45mm at 400X and 0.18mm at 1000X.  See micrometer.
冠心病能治好吗Fixed Arm:  A type of stand ud with low power microscopes.  The arm and body are integral parts of the microscope and connected solidly to the ba. 
Focus:  A means of moving the specimen clor or further away from the objective lens to render a sharp image.  On some microscopes, the stage moves and on others, the tube moves.  Rack and pinion focusing is the most popular and durable type.
Head:  The upper part of the microscope that contains the eyepiece tube and prisms.  A monocular head has one eyepiece, a binocular has two (one for each eye), a dual head has two but they are not together, and a trinocular head has three, one which is generally ud for a camera connection.
Illuminator:  A light source mounted under the stage.  Three types of light are commonly ud:  Tungsten, Fluorescent and Halogen.  Tungsten is the least expensive and most common.  Fluorescent is bright, white and runs cool and Halogen is very bright and white but gives off heat like tungsten.
 Immersion Oil:  A special oil ud in microscopy with only the 100X objective lens (usually at 1000X total power).  A drop is placed upon the cover slip and the objective is lowered until it just touches the drop.  Once brought into focus, the oil acts as a bridge between the glass slide and the glass in the lens.  This concentrates the light path and increasing the resolution of the image.  Both Type A and Type B are commonly ud in light microscopy and the only difference is the viscosity (B is more viscous).
Inclination Joint:  Where the arm connects to the ba, there may be a pin.  If so, you can place one hand on the ba and with the other grab the arm and rotate it back.  It will tilt your microscope back for more comfortable viewing.  One drawback of tilting it back is that wet samples will run off the slide.
Interpupiliary Adjustment (also called Inter Ocular Distance I.O.D.):  When using a stereo or binocular microscope there must be an adjustment for the distance between the viewer’s eyes.  A young child will have a small interpupiliary distance and an adult a larger one.  The eyepiece lens will spread apart or get clor together to fit each individual.  This should be the first adjustment to make so that you are comfortably viewing the specimen with both eyes.
Mechanical Stage:  A mechanical way to move the slide around on your stage.  It consists of a slide holder and two knobs.  Turn one knob and the slide moves toward or away from you.  Turn the other knob and the slide moves left and right.  Since everything is upside down on a (high power) microscope it takes some getting ud to but it is very convenient to have one especially when obrving moving specimens like protozoans or other pond water critters.  Microscopes either have the bolt on mechanical stage that can be added (to many models) at any time or the integral mechanical stage that comes built in to the microscope. 
Micrometer:  Also called a micron it is the metric linear measurement ud in microscopy.  There are 1000 microns in a millimeter.  If something is 1.8mm long then it can also be expresd as 1,800 microns (or micrometers) long.
Mirror:  Allows you to direct ambient light up through the hole in the stage and illuminate the specimen.
Monocular Head:  A microscope head with a single eyepiece lens.
Nopiece:  The part of the microscope that holds the objective lens also called a revolving nopiece or turret.
Numerical Aperture (N.A.):  This is a number that express the ability of a lens to resolve fine detail in an object being obrved.  It is derived by a complex mathematical formula and is related to the angular aperture of the lens and the index of refraction of the medium found between the lens and the specimen.  To get the best possible image, you should have a condenr system that matches or exceeds the N.A. of the highest power objective lens on your microscope.  (note, N.A. is only important with high power microscopes). 
Objective Lens:  The lens clost to the object.  In a stereo (low power) microscope there are objective pairs, one lens for each eyepiece lens.  This gives the 3-D effect.  On a high power binocular model there is still only one objective lens so no stereo vision.
Oil Immersion Lens:  An objective lens (usually 100X or greater) designed to work with a drop of special oil placed between it and the slide.  With oil, an increa in resolution will be noticed.    Also, e "Immersion Oil" above.
Parcentered:  This is an alignment issue.  When changing from one objective lens to another, the image of the object should stay centered.  Test this by centering something in your field of view.  Change to a higher power.  Is it still centered?  Almost all  microscopes are parcentered.
Parfocal:  This is a focus issue.  When changing from one objective to another, the new image should be either in focus or clo enough so that you can refocus with only minor adjustments.  Most microscopes are parfocal.
Pointer:  When you look through the eyepiece lens, you may e a pointer.  By turning the eyepiece, you can rotate the pointer around. 
Post Stand:  A type of stand ud with low power microscopes.  It consists of a single post rising vertically from the ba.  The microscope body can rotate about the post and also be moved up and down on it. 
Rack and Pinion:  牡丹开花时间The rack is a track with teeth and the pinion is a gear that rides on the teeth.  By turning a knob, the pinion gear moves along the rack.  The systems are ud in focusing mechanisms, in Abbe condenr focusing systems, and on mechanical stages to move the slide around.
Rack Stop (or Safety Rack Stop):  Usually t at the factory, the rack stop keeps you from cranking the objective lens too far down (damaging something).  If you are using a very thin slide, you may find that you can't get the high power objective lens clo enough to the slide to focus.  Here you can either adjust the rack stop or place a thin glass slide under your original slide, making it clor to the lens.
Resolution:  The ability of a lens system to show fine details of the object being obrved.
Reticle:  A very tiny grid pattern inrted in an eyepiece lens.  It is ud to make actual measurements of the size of objects en through the microscope.
Revolving Nopiece:  See nopiece
Ring Light:  An independent light that usually connects to the microscope body and gives off a ring of light. 
Semi-Plan Lens:  Lens are never perfect.  If you were looking at something perfectly flat, you might find that much of the center part of your field of view is in focus but out on the edges it is fuzzy and a bit out of focus.  Semi-plan lens improve this deficiency by showing sharper images and less aberrations in the perimeter of the field of view.  They are better than standard achromatic lens but cost quite a bit more.
五香花生
Slide:  A flat glass or plastic rectangular plate that the specimen is placed on.  It may have a depression or well to hold a few drops of liquid.
Slip Clutch:  When students bring the focus all the way up or down and continue to try turning the knob, damage to the focusing system can occur if there wasn't a slip clutch.  It is a mechanical device that protects the gears of the microscope.
Stage:  The flat plate where the slides are placed for obrvation.
Stage Clips:  Clips on the stage ud to hold the slide in place.
Stage Plate:  On a low power microscope, there is a frosted circular glass plate that fits in over the lower illuminator.  This is called the stage plate.  See also contrast plate.
Stand:  On a low power microscope, the type of connection between the microscope body and the ba.  There are three main types:  the post, the fixed arm and the universal boom stand. 
Stereo:  Related to microscopes, eing with both eyes through parate eyepiece and objective lens.  With two objectives, the image looks 3-D, we e it in "stereo"!  See also Binocular head.
Student Proofed:  We're always trying to outwit the students.  Many of the classroom type microscopes have just about everything locked down.  You need special tools to remove eyepiece lens, objective lens and they have all the safety devices like the rack stop.  Not totally student proofed (like drop proof!) but clo.
中国存款利率
Sub-stage:  The area below the stage as in "sub stage illuminator"
T-mount:  A type of adapter ud to mate still cameras (usually 35mm) to microscopes
家常炸丸子
Tension Adjustment:  This is an adjustment of the focusing mechanism that is made at the factory.  It is t so that the instrument is easy to focus but also tight enough so that the stage doesn't drift when you are not focusing.  Stage drift is caud by the weight of the stage (or tube) automatically unfocusing the microscope.
Trinocular Head:  Available on both high and low power microscopes, tri heads have two eyepiece lens (one for each eye) and a third port at the top for a camera.  Some microscopes give you the option of nding all the light to the tri port, or perhaps half and half, or maybe 70/30%.  On some stereo tri heads with dual power,  the tri port transmits the image through the t of lens not being ud by the stereo eyepieces.
Turret:  See nopiece.
Universal Stand:  A long boom type arm ud to support a (low power) microscope body.  It has many adjustments allowing the microscope to be aligned in a wide variety of configurations.  Generally one us an external (like a fiber optic) light source with a universal stand.
Widefield eyepiece lens:  The are wide diameter glass eyepiece lens.  They offer the greatest field of view when looking at specimens. 
X:  Times as in 200X or two hundred times magnification.  The magnification of a microscope is determined by multiplying the power of the eyepiece lens by the power of the corresponding objective lens.
XR:  The X is times (e above) and the R stands for retractable.  The objective lens have a spring loaded tip so if they hit the slide, they will retract, and telescope inward.  This prevents damage to the lens or slide