Notice
Thank you for using our product. Plea read this ur manual before operating the product.
This ur manual is written according to MDD93/42/EEC standard and instruction of EUROPEAN UNION. The content of this manual is accord with the product. If the software upgrade or modified, this manual will change without notice.
This ur manual describes speciality and request of this product, main configuration, capability, specification, transportation, installation, operation, maintenance, stock method and safety precaution for operator and equipment. Detailed content plea refer every chapter.
Plea read this ur manual before using the product, and operate according to this manual. This manual will show you the operation step, the wrong operation, which will lead dangerous. As for the dangerous caud by wrong operation, which violates the rules in manual, our company without responsibility for this kind of danger and will not maintain free for this kind of malfunction.
Warning:
1 plea check patie nt’s skin carefully, don’t emplace the equipment on dropsical or
fragility part.
2 Plea don’t look the radiation parts of apparatus directly(infrared ray is invisible), even
maintainer should not watch it, too. It will be hurtfully for people’s eye.
3 Th e testee’s fingernail should not daub cosmetic, for example: enamel.
4 The testee’s fingernail should not be too long.
5 As for detailed content of clinic restrict and contraindication, plea refer to relevant
medicinal literature.
6 this equipment can be ud for diagno, but not therapy.
7 plea don’t u this equipment in combustible environment.
8means “Attention”, plea refer to enclod file(this ur manual)
This ur manual is written by Contec Medical Systems Co., Ltd. The copyright belongs to Contec Medical Systems co., Ltd.
Catalogue
1 SAFETY ................................................................................................................................................... 1
2 BRIEF INTRODUCTION ...................................................................................................................... 2
3 PRINCIPLE AND ATTENTION ITEMS ............................................................................................. 2
4 TECHNICAL PARAMETER ................................................................................................................ 4
5 PRODUCT ACCESSORY ...................................................................................................................... 4
6 INSTALLATION ..................................................................................................................................... 4
7 OPERATION INTRODUCTION .......................................................................................................... 9
8 MAINTENANCE AND PRESERV ATION .............................................................. 错误!未定义书签。
9 POSSIBLE PROBLEMS AND SILUTIONS .................................................................................... 15
10 15
11 PRODUCT DIMENSION ................................................................................................................. 16
1 Safety
1.1 Introduction of safe operation9大行星
美容小秘方✧Periodic examination can make sure the equipment is not damaged obviously, which may affect
safty and monitor capability. We recommend Periodic examination every week. If there is obviously damage, plea stop using the equipment.
✧Only the qualified maintainer appointed by our company can maintain this equipment, plea
don’t maintain it by yourlf.
✧This equipment is not compatible with other equipments except the appointed equipments in this
ur manual.
1.2 Warning
四级听力题型
Danger of explosive: plea keep this equipment away from the tinder such as anaesthetic.
Plea don’t u the pul oximeter with Magnetic resonance imaging(MRI)and CT examination.  As for discard as uless of the equipment and accessories(include battery), plea follow the local law and regulation.
1.3 Attention
The operation environment should avoid dust, shake, corrosive and combustible, the temperature should not be too high or too low
If this equipment is dabbled or clotted by water, plea stop operation.
Plea don’t operate the equipment at once when it is taken from cold environment to warm moist environment.
Plea don’t operate the front panel switch with edgily tool.
幼儿发热Plea don’t u high temperature high pressure or gas disinfection on this equipment. The operation for cleanout and disinfection plea refer to ur manual.
Plea don’t immerge the equipment in liquid, when cleaning the equipment with disinfector, plea don’t spill the liquid on equipment.
The temperature of water for cleaning the equipment should under 60℃.
To some patient, who finger is too thin or too cold, the measure may not be normally, under this condition, plea nip the equipment onto thick finger(thumb or middle finger), and make sure the patient’s finger inrt deep enough.
This equipment can be ud for children more than 4years old and adult (weight between 15-110 kilogram).
成都武侯祠
This equipment can be ud for children more than 4years old and adult (weight between 15-110 kilogram).
If there is something abnormal during testing, plea pull your finger out, then inrt again, the test will resume normal.
The usage life-span of the equipment is 5 years since first electrified usage.
This product doesn’t include function of alarm when the battery with low electricity volume, this equipment only show battery volume, so plea change battery in time when battery with low electricity volume.
This equipment with function of alarm for oxemia and pul rate. When the appointed measure result over tted normal parameter range, it will alarm.
2 Brief Introduction
SPO2 is the percentage of HbO2 in Hb, which is an important physiological parameter for respiration system. To make sure measuring SPO2 accurate and convenient. Our company develop the pul oxemeter, this equipment can measure pul at the same time.
The oxmeter made in Contec with small bulk, low consume of energy, convenient to schlep. When you inrt finger into the oximeter probe, the SPO2 parameter will be display in the screen.
2.1 Product Properties
1 Operation of the product is simple and convenient.
2 The product is small in volume, light in weight and convenient in carrying with printing function.
3 Power consumption of the product is low and the originally-equipped rechargeable Li-battery can be operated continuously for printing for 1 hour.
4 Power display
2.2 Product categories: II b (e MDD appendix IX, No. 10)社区公益活动
2.3 Proleptic u
This oximeter can detect SPO2 and Pul Rate through finger. It can be ud for family, hospital(clinic u, eg. internal medicine&surgical, anaesthesia, paediatrics, emergency treatment, etc.), oxygen bar, community medical treatment and athletics care(can be ud before or after sports, but not during sports)
This product can be ud for monitoring patient continuously.
2.4 Environment requirement
Ambient temperature 10℃~40℃
Prerving temperature -20℃~55℃
Ambient humidity working humidity: 35%~80%
prerving humidity: 10%~100%
Barometric scope working scope: 70kPa~106kPa
幼儿园足球游戏prerving scope: 50kPa~106kPa
3 Principle and Attention Items
3.1 Work Principle
The working principle of oxemeter is according to the absorb spectrum characteristic of R Hb and O2 Hb under red glow. According to Lambert Beer law, we get formula for data processing. Its working principle adopt photoelectricity oxemeter examine technology and dimension pul sign technology, which take two beams of light with different wavelength (660nm red glow and 940nm sub infrared ray) irradiate people’s finger nail edge to get measure signal data with clairvoyant nip nsor. The acquired data be procesd in electrocircuit and microprocessor(MPU) and then display the result on LTD.
Diagram of work principle
1. Glow and Infrared-ray Emission Tube
2. Glow and Infrared-ray Receipt Tube
3.2 Attention Items of operation
1 Plea inrt your finger according to this manual and attached drawing, or the measured result may
弟弟用英语怎么说
be wrong.
2. there is testee’s little artery between photoelectricity receiver tube and irradiation tube.
3. plea avoid measure the part, which with artery canal, blood pressure strap or mainline.
4. plea avoid adhesive plaster on the way of light.
5. if the environment with too strong light, it will affect measure result, besides fluorescence lamp,
infrared ray heater, point-blank sunshine.
6. testee’s movement and electricity surgery equipment will influence the precision of measure result.
7. the testee’s fingernail mustn’t with nail polish and other cosmetic.
3.3 Clinical Restriction
1 As the measure is taken on the basis of arteriole pul, substantial pulsating blood flow of subject is required. For a subject with weak pul due to shock, low ambient/body temperature, major bleeding, or u of vascular contracting drug, the SpO
2 waveform (PLETH) will decrea. In this ca, the measurement will be more nsitive to interference.
2 For tho with a substantial amount of staining dilution drug (such as methylene blue, indigo green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb) or thiosalicylic hemoglobin, and some with icterus problem, the SpO2 determination by this monitor may be inaccurate.
3 The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for rious error of SpO2 measure.
4 As the SpO2 value rves as a reference value for judgement of anemic anoxia and toxic anoxia, some patients with rious anemia may also report good SpO2 measurement.