Technical requirements for medical device product registration number:
Disposable Sterile PTCA balloon catheter
1、Model / Specifications and divide instructions
1、1 Model
PTCA-XX-XX
Balloon Working Length mm
Balloon nominal Diameter mm
Product Name Code
Model mark example:PTCA—2.0—10 reprents balloon nominal outer diameter is 2.0mm and working length is 10mm of disposable Sterile PTCA balloon catheter
1、2 Specifications 
                                    PTCA balloon catheter specification list
Specifications
OD
Length
The effective length of the catheter
Max guide wire diameter
Nominal pressure
RBP
DK-PTCA-1.5-10
1.50mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-1.5-20
1.50mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-1.5-30
1.50mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-1.75-10
1.75mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-1.75-20
1.75mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-1.75-30
1.75mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.0-10
2.00mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.0-20
2.00mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.0-30
2.00mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.5-10
2.50mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.5-20
2.50mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.5-30
2.50mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.75-10
2.75mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.75-20
2.75mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-2.75-30
2.75mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-3.0-10
3.00mm
10mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-3.0-20
3.00mm
20mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-3.0-30
3.00mm
30mm
142cm
≦0.014in
8ATM
16ATM
DK-PTCA-3.5-10
3.50mm
10mm
142cm
≦0.014in
6ATM
16ATM
DK-PTCA-3.5-20
3.50mm
20mm
142cm
≦0.014in
6ATM
16ATM
DK-PTCA-3.5-30
3.50mm
30mm
142cm
≦0.014in
6ATM
16ATM
DK-PTCA-4.0-10
4.00mm
10mm
142cm
≦0.014in
6ATM
16ATM
DK-PTCA-4.0-20
4.00mm
20mm
142cm
≦0.014in
6ATM
16ATM
DK-PTCA-4.0-30
4.00mm
30mm
142cm
≦0.014in
6ATM
16ATM
1、3 Divide instructions
Disposable Sterile PTCA balloon catheter consists of balloon, connecting catheter, tip, support catheter and Handle. According to different balloon diameter to divide the balloon into different specifications and sizes. According to medical device classification rules 6877, the balloon catheter belonging to Class Ⅲ interventional equipment intravascular catheter.
2、Performance
2、1 Appearance
As with normal vision or corrected vision at a magnification of 2.5 times the conditions of inspection, the effective length of the outer catheter should be clean and free of impurities. There should be no processing defects and surface defects on the surface of catheter effective length and end.
The inner and outer surface of the distal end components of the catheter should have silicone oil (polydimethylsiloxane) lubricant coating. With normal vision or corrected vision inspection, the outer surface of the catheter should not e the convergence of lubricant droplets.
Note:The inner and outer surface of the distal end components of the catheter is the out surface from the tip (figure 1) to the transition weld (figure 7),and the inner surface from guide wire lumen (from the tip (Figure 1 below) to quickly switch ports (figure 5)).
2、2 Size requirements
2.2.1 Balloon catheter size requirements reference table1, specification list reference table 2.
1、Tip  2、Balloon 3、Radiopaque ring  4、Out catheter 5、Quickly switch ports  6、Guide wire  7、Transition weld 8、Hypo Tube(FEP Coated)  9、Bending resistant tube  10、Handle  11 Balloon effective length(    L)  12、catheter effective length
                              Pic 1. PTCA balloon catheter schematic
Pic 2.  Tip schematic
Table 1 disposable Sterile PTCA balloon catheter size and tolerance requirements
Name
Unit
Standard
Tolerance
Remark
Balloon  diameter under nominal pressure
mm
D
±0.125
Balloon effective length
mm
L
±10%
Asmbly integrated tolerances±10%,
Reference pic 1.
Radiopacity Length(Platinum ring spacing)
mm
(L-1)mm(L is balloon effective length)
±10%
Adapted to the maximum diameter of the guide wire
Inch
NA
NA
0.014in Guide wire pass through guide wire lumen smoothly(Wear into the tip,piercing out from  fast switching port 5)
Max catheter OD in components
mm
1.0
±0.1
Proximal welding point
Catheter effective length
cm
142
±2
Reference pic 1.
Tip size
mm
2
±0.5
Reference pic 2. outside of crater
2.2.2 Tapered connector
Connector of balloon catheter must be tapered connector. Size should be consistent with GB / T 1962.1-2001 requirements.