PP-031 Qualification of the Repeater (Baxa) pump in a centralid cytotoxics preparation unit
意大利语在线课程
big data期刊名称: European Journal of Hospital Pharmacy
上海英语枪手作者: Schorgmeier, A, Michelet, V, Faure, P, Jourdan, N
年份: 2014年
启发英文期号: 第Suppl 1期
reciprocally摘要:BackgroundOur centralid cytotoxic preparation unit generates over 50,000 preparations per year, of which about 4% are standardid rounded do preparations performed manually. In order to optimi the production and to increa volume and quality of the preparations, the unit acquired a mi-automatic Repeater pump.PurpoTo test the analytical performance of the pump by establishing a qualification protocol in order to integrate it in the current functioning of the unit.Materials and methodsThe parameters assd were linearity, precision (repeatability and reproducibility) and accuracy. To this end, transfers of solvent (5% gluco and 0.9% sodium chloride) were performed for
different volumes: 1, 10, 20, 50, 100, 150, 250, 500 and 900 mL (which is the maximum volume ud). Each volume was repeated six times