Update to BleaFocus
Equipment affected: BleaFocus
Details: The BleaFocus has had a major update. Most of the updates are cosmetic电动燃油泵
上海英语口语班
however, the ventilator has been more cloly integrated and has required some additional internal wiring and modification of internal tubing. The pneumatic module has also been updated. The ventilator will be in position when shipped.
The details will be added to the rvice manuals in the next updates.
Wires routed internally
Shown in red
Back Cover
Wires routed Original Focus
Focus Facelift
TECHNICAL BULLETIN
BUSINESS QUALITY DOCUMENT
im
Changes:
Ventilator is now turned on / off by the main on / off switch on the front of the BleaFocus, not by the Trakwheel. This has necessitated new vent definition files the will end in the letter F (750F/950F etc). The absorber and bellows connections are now in the upright by the absorber mounting bracket.
The O2 cell connection is under the front of the work surface.
The patient flow nsor connection is now on the front corner by the
absorber.
There is no AIR / N2O switch.
The safety valve is now fixed at 3 to 4PSI The O2 flush is fixed at 45 to 55 lpm
The pneumatic block has been updated, to include the condary
regulators.
The ventilator is clod in at the rear.
The ventilator gas supply is now a 6mm push fit connector in place of the NIST/DISS fitting ud in the past. If no ventilator is order then a DISS or mini Schrader valve will be fitted.
互联网创业趋势The Fresh gas ho will be blue silicon not black rubber.
Absorber bag/vent switch cable connects to new socket under the work surface near to the absorber mounting plate.
The 17mm drive gas ho connects to a new connector in the upright
TECHNICAL BULLETIN
gleee
BUSINESS QUALITY DOCUMENT
O2 cell cable connection is under the work surface near the front corner. Patient flow nsor tube connection, on the front corner near the absorber.
On/Off switch also turns the ventilator On/Off. No Air/N2O switching. Flush fixed at 45 – 55 LPM. Safety valve fixed at 3-4psi.
TECHNICAL BULLETIN
BUSINESS QUALITY DOCUMENT
New pneumatic layout
开心就好英文CGO
TECHNICAL BULLETIN
headaches
BUSINESS QUALITY DOCUMENT
深圳英孚教育
TECHNICAL BULLETIN
hymnBUSINESS QUALITY DOCUMENT
Note This BleaFocus tubing is using ANSI colour scheme where:- O2 = Green, Air = Yellow, N2O = Blue For ISO systems O2 = White, Air = Black, N2O = Blue